Dimension ICON Bruker
-----------------------------------------------------------
Note: problem laser -- do not scan more than 20 x 20 µm² including offsets
Do not use z sensor as topographical information
Horizontal deflection blocked
---------------------------------------------------
Sample size: 20 centimeter diameter
Scan range: 100µm x 100µm, Z range :10µm
X and Y piezo linearization Feedback (close loop)
Lateral resolution: nanometric, Vertical resolution : 30pm
Working mode ; AFM Tapping, AFM Peakforce, EFM, KPFM, CAFM, PeakForce TUNA, PFM, SThM, Force spectroscopy, SCM, SGM
Environment: Ambient air, Nitrogen gas and Liquid
Temperature working range: -25°C-150°C
Term of use: Free Access